Application
This product is suitable for digestion, acid removal and pretreatment in environmental protection, disease control, scientific research and quality inspection.
Characteristics
1.Three-dimensional surrounding heating method ensures a uniform temperature; special thermal insulation materials ensure energy-efficient heating.
2.Resistant to corrosion, high temperature, acid and alkali. Easy to clean.
3.Large heating area enables the instrument to handle multiple samples at the same time. High temperature uniformity between holes ensures high digestion rate.
4.Split type anticorrosive design of the control part and heating template prevents the acid erosion to the instrument.
5.PID parameter self-check function and adjustable heating rate. Temperature fluctuation of the single hole is ±1.0℃.
6.Program LCD ,30 program control, working time,temperature gradient, and rate free setting achieved intelligent digestion
7.Normal pressure digestion mode and design of recirculation hole can save reagents supplies about 50%.
8.Featured by "One-stop" digestion concept. Namely, digestion, acid removal and the constant volume can be completed in the same digestion tank.
9.Over-temperature alarm ensures the safety of the experiment.
Typical Applications
EPA method of soil treatment : 3010a、3010b、3050b、3060
EPA method of water treatment: 200.2、200.7、200.8、200.9、245.1、365.1
Electronic product treatment method: SJ/11365-2006
GB method of paint treatment: GB/T22788-2008
GB method of soil treatment: GB/T22105.1-2008 GB/T22105.2-2008 GB/T22105.3-2008 GB method of foodstuff: GB/T17138-1997 GB/T5009.11-2003 GB/T5009.17-1996
All wet digestion: user-defined programs
Main Technical Parameters
Model | XJT20-20 | XJT40-20 | XJS60-24 | XJS20-42 | XJS40-42 |
Power voltage(V) | 220V/50Hz | ||||
Heating material | High purity graphite+Teflon | High purity Graphite | |||
Heating power (W) | 1800 | 2100 | 3000 | 3000 | 3000 |
Highest Temp erature (℃) | 240 | 450 | |||
Temp. resolution(℃) | 0.1 | ||||
Temp. control precision(℃) | ±1 | ||||
Temp. uniformity between holes(℃) | ±1.5 | ||||
Sample capacity(ml) | 20x50 | 40x50 | 60x50 | 20x50 | 40x50 |
Digestion hole size(D╳H)mm | Φ31x50 | ||||
Outline dimensions of heating part | 301x257x180 | 413x290x180 | 494x331x180 | 331x287x180 | 443x320x180 |
Outline Dimensions(W×D×H) mm | 220x170x77 |