ZHPZ-115 / ZQPZ-115 Combined Shaking Incubator
The products are widely used in microbial, viral, bacterial culture, fermentation, hybridization,
biochemical reaction and cell tissue research. For temperature, oscillation frequency, amplitude have different requirements of culture and enzyme engineering research to provide effective help in medicine, biology, molecular science, pharmaceutical, food, environmental protection and other research applications have a wide range of important roles
Product Features
Humanized design
1.Using mirror stainless steel integrated forming studio, each corner arc is excessive without dead corners, no welding gaps, can be a full range of dead corner washing, waterproof to prevent bacteria growth, easy to clean.
2.Maximum three-layer superposition combination method occupies a small space and can effectively use the laboratory space. Each layer is independently controlled to set different temperature and speed.
3.Pull-out quick loading pallet is convenient for loading and unloading and replacement of shaking plate flask.
4.Hollow tempered glass door, easy to observe the internal situation; The side is equipped with a ¢30mm test hole, which is convenient for testing and meeting the sample's demand for oxygen.
5.The cooling system is designed with a dust filter for easy cleaning and dust removal
Unique performance
1.Imported fluorine-free refrigeration compressor, with a unique refrigeration working mode, to ensure that the compressor will not shut down due to its own overheating, so as to achieve high-quality temperature control effect.
2.Centrifugal circulation duct, forcing the air flow in the working room, the temperature is more uniform
3.Endless adjustment of circulating wind speed in working room.
4.High quality servo motor control scheme, accurate revolution measurement, long life, maintenance-free.
5.Can carry out positive and negative work setting, timing, fixed value programmed operation.
6.With door control function, the door automatically stops turning.
7.It is driven by eccentric wheel shaft with three axes, which has stable operation and long service life.
8.Microcomputer large screen temperature control display, setting, observation clear and intuitive.
9.LCD large screen LCD display and true color touch screen programmed temperature control system can be selected to fully meet your needs (true color touch screen operating system can be optional)
10.With timing function, the time setting range is 0~9999 minutes (hours).
Safety guarantee function
1.With power failure memory function, to avoid data loss caused by power failure and crash.
2.The control system integrates lighting and UV sterilization for easy observation and sterilization.
3.With over temperature sound and light alarm, automatic power off function. It has the function of fault self-diagnosis.
4.With door control function, the door automatically stops turning.
Optional function
1. Micro printer can print data regularly according to needs
2. 485 Interface remote control system: remote control and data analysis and export according to work needs .
3. The remote monitoring function of the mobile APP client can be upgraded.
Main Technical Parameters
Model | ZHPZ-115 | ZQPZ-115 |
Volume(L) | 150 | |
Intelligent control system | LCD(Liquid crystal display screen) | |
Shaking frequency(rpm) | 10-350 | |
Shaking frequency precision(rpm) | ±1 | |
Orbit(mm) | Ф26 | |
Temp.variation(℃) | RT +5~65 | 4~65 |
Temp. resolution(℃) | 0.1 | |
Temp. uniformity(℃) | ±0.5(37℃) | |
Platform size (mm) | 580×460 | |
Max.capacity/layer(ml×piece) | 100×53 or 250×30 or 500×20 or 1000×12 or 2000×8 or 3000×5 | |
Control mode | P.I.D Microcomputer intelligent control | |
Timing Range | 0~9999 min(hour) | |
Power/layer(Kw) | 0.5 | 0.7 |
Power supply(V) | AC220±10% / 50~60Hz | |
Outline dimensions/layer(mm) | 950×800×600 |
Tagging:
1.The performance parameter test under no-load conditions is: ambient temperature of 20 ℃, ambient humidity of 50% RH
2.Working environment temperature: 5-25 ℃
3.To fully utilize the effectiveness of the equipment, the fixture configuration for all specifications of products leaving the factory shall be based on the standard capacity configuration.